MTS-1450 Meter Calibration System

Highlights

  • Calibrate voltage, current, Watt, VAr and frequency meters
  • Calibrate all single and three phase power meters
  • Semi-automated or fully automated testing capable


Specifications


Power Supply
  • Single phase 108-133 VAC/60Hz (or 216-264 VAC/50Hz), 200VA max, factory set

Outputs

Single phase AC voltage:

  • 0-155V rms, 15 VA maximum
  • Accuracy +/-0.1%of full scale +/-2 least significant digits, for 10%-100% of full scale
  • Four color coded output terminals, A, B, C, N provided.
  • Selectable phase-phase (A-B, C-B, C-A) or phase-neutral (A-N, B-N, C-N) output
  • All phase-neutral, or A-B plus C-B outputs, may be selected simultaneously


Single phase AC current:

  • 0-6A rms, 18 VA maximum
  • Accuracy +/-0.1%of full scale +/-2 least significant digits, for 10%-100% of full scale
  • Three pairs of color coded output terminals, A, B, C provided.
  • Each phase output and polarity independently controllable, and group reverse switch provided


Phase angle:

  • Power factor is pushbutton selectable between 1.0 and 0.5
  • Angle between current and voltage adjustable between -90 to +90 deg, accuracy resolution: 1 degree.


Frequency:

  • Default 60.00Hz
  • Variable 50.00 - 70.00 Hz (crystal controlled, 0.01% accuracy)
  • Optional 45.00 - 55.00 Hz range


RS-232C Interface

  • Standard DB-25 female connector
  • Baud rates: 300 to 19200 baud
  • Easy-to-use terminal mode
  • Interface is optically isolated from power circuitry

Additional Features

Stepping:

  • Voltage, current, Watts, VARs, or frequency may be stepped
  • Start, stop, step value, and delay between steps independently adjustable


Configuration:

  • Multiple configuration setups with custom procedure name, voltage current output selection, P.T., C.T. ratios, and user defined prompts can be saved in non-volatile memory.
  • Primary readings can be displayed


Default values:

  • Power-up default values of all output parameters plus primary P.T. C.T. ratios can be saved in non-volatile memory.


Physical Characteristics

  • 14.5"W x 7.5"H x 15.5"D (36.8cm W x 19cm H x 39.4cm D)
  • Weight: 24.0 lbs (11 kg)


Options

  • Option 02: Cordura carry case
  • Option 03: Test leads- 6ft leads for connecting the voltage and current outputs to the meter(s) under test
  • Option 04A: 120V nominal AC line voltage input
  • Option 04B: 240V nominal AC line voltage input
  • Option 05A: 60Hz nominal AC frequency
  • Option 05B: 50Hz nominal AC frequency
  • Option 07: Extra manual
  • Option 08: 1 Year extended warranty. Additional year for a total of 2 years
  • Full-Service Agreement Preventative maintenance, calibration and service plan

NOTE: All specifications are subject to change