MTS-1400 Transducer Test System

Highlights

  • Calibrate voltage, current, Watt, VAr and frequency transducers
  • Calibrate all single and three phase power transducers
  • Direct display of percent error
  • Connect up to 9 transducers simultaneously
  • Semi-automated or fully automated testing capable

For calibration of meters only, see our MTS-1450 Meter Calibration System


Specifications


Power Supply
  • Single phase 108-133 VAC/60Hz (or 216-264 VAC/50Hz), 200VA max, factory set

Outputs

Single phase AC voltage:

  • 0-155V rms, 15 VA maximum
  • Accuracy +/-0.1%of full scale +/-2 least significant digits, for 10%-100% of full scale
  • Four color coded output terminals, A, B, C, N provided.
  • Selectable phase-phase (A-B, C-B, C-A) or phase-neutral (A-N, B-N, C-N) output
  • All phase-neutral, or A-B plus C-B outputs, may be selected simultaneously


Single phase AC current:

  • 0-6A rms, 18 VA maximum
  • Accuracy +/-0.1%of full scale +/-2 least significant digits, for 10%-100% of full scale
  • Three pairs of color coded output terminals, A, B, C provided.
  • Each phase output and polarity independently controllable, and group reverse switch provided


Phase angle:

  • Power factor is pushbutton selectable between 1.0 and 0.5
  • Angle between current and voltage adjustable between -90 to +90 deg, accuracy resolution: 1 degree.


Frequency:

  • Default 60.00Hz
  • Variable 50.00 - 70.00 Hz (crystal controlled, 0.01% accuracy)
  • Optional 45.00 - 55.00 Hz range


Transducer Inputs

  • 9 input channels for DC current from external transducers
  • 20 mA continuous, 30mA absolute maximum
  • 0-1mA, (4-20mA optional) ranges
  • Full-scale inputs correspond to 150V, 5A, 500W, 500VARs per phase. Other full scale ranges [100-155V, 1.5-5.5A, 100-550 W/VAR] programmable (at reduced accuracy). Other full scale inputs available via special order.
  • Frequency transducer full scale input programmable to any frequency between between 50.00 - 70.00Hz, unipolar or bipolar.
  • 1, 2, or 3 element power transducers accepted. Special scaling for 2.5 element transducers selected via menu (1mA range only).
  • Quick connect cable connector provided for channels #1-8, five-way binding posts for channel #9
  • Load on selected channel adjustable (70 ohms - 100 kohms +/-1% +/-5ohms in 120 ohms steps.


%Error Measurement

  • Error between device under test and internal reference standard is displayed directly in percent of full scale for volts, amps, Watts and VARs. [Bracketed values apply to menu programmed non-standard output ranges]
  • Range: 0.01%-9.99% [0.1%-9.9%]
  • Accuracy: +/-0.1% [+/-0.2%] (volts, amps, Watts), or +/-0.125% [+/-0.25%] VARs, for values of 10%-100% of full scale and load resistance > 5% of maximum value.
  • Rated accuracy maintained for load resistances 5-100% of ( 10kohm (0-1mA range), 700ohm (4-20mA range) ).
  • Frequency error measurement: +/-0.135% of full scale in mA
  • Minimum 20 minutes warmup time required to reach rated accuracy, 30 minutes if initial instrument temperature is less than 18 deg. C
  • Alarm limit programmable 0.10%-2.00%
  • External milliamp measurement accuracy: +/-0.125% of full scale


RS-232C Interface

  • Standard DB-25 female connector
  • Baud rates: 300 to 19200 baud
  • Easy-to-use terminal mode
  • Interface is optically isolated from power circuitry

Additional Features

Stepping:

  • Voltage, current, Watts, VARs, or frequency may be stepped
  • Start, stop, step value, and delay between steps independently adjustable


Configuration:

  • Multiple configuration setups with custom procedure name, voltage current output selection, reference input selection, P.T., C.T. ratios, and user defined prompts can be saved in non-volatile memory.
  • Primary readings can be displayed


Default values:

  • Power-up default values of all output parameters plus primary P.T. C.T. ratios can be saved in non-volatile memory.


Physical Characteristics

  • 14.5"W x 7.5"H x 15.5"D (36.8cm W x 19cm H x 39.4cm D)
  • Weight: 25.0 lbs (11 kg)


Options

  • Option 02: Cordura carry case
  • Option 03: Test leads- 6ft leads for connecting the voltage and current outputs to the transducer(s) under test
  • Option 04A: 120V nominal AC line voltage input
  • Option 04B: 240V nominal AC line voltage input
  • Option 05A: 60Hz nominal AC frequency
  • Option 05B: 50Hz nominal AC frequency
  • Option 06: 4-20mA range
  • Option 07: Extra manual
  • Option 08: 1 Year extended warranty. Additional year for a total of 2 years
  • Full-Service Agreement Preventative maintenance, calibration and service plan

Download full MTS-1400 Series brochure 148kB PDF format

NOTE: All specifications are subject to change