- Calibrate voltage, current, Watt, VAr and frequency transducers
- Calibrate all single and three phase power transducers
- Direct display of percent error
- Connect up to 9 transducers simultaneously
- Semi-automated or fully automated testing capable
|
|
|
For calibration of meters only, see our MTS-1450 Meter Calibration System
|
|
|
| Power Supply |
- Single phase 108-133 VAC/60Hz (or 216-264 VAC/50Hz),
200VA max, factory set
|
| Single phase AC voltage: |
- 0-155V rms, 15 VA maximum
- Accuracy +/-0.1%of full scale +/-2 least significant digits,
for 10%-100% of full scale
- Four color coded output terminals, A, B, C, N provided.
- Selectable phase-phase (A-B, C-B, C-A) or phase-neutral
(A-N, B-N, C-N) output
- All phase-neutral, or A-B plus C-B outputs, may be selected
simultaneously
|
|
Single phase AC current:
|
- 0-6A rms, 18 VA maximum
- Accuracy +/-0.1%of full scale +/-2 least significant digits,
for 10%-100% of full scale
- Three pairs of color coded output terminals, A, B, C provided.
- Each phase output and polarity independently controllable, and
group reverse switch provided
|
|
Phase angle:
|
- Power factor is pushbutton selectable between 1.0 and
0.5
- Angle between current and voltage adjustable between -90
to +90 deg, accuracy resolution: 1 degree.
|
|
Frequency:
|
- Default 60.00Hz
- Variable 50.00 - 70.00 Hz (crystal controlled, 0.01% accuracy)
- Optional 45.00 - 55.00 Hz range
|
|
Transducer Inputs
|
- 9 input channels for DC current from external transducers
- 20 mA continuous, 30mA absolute maximum
- 0-1mA, (4-20mA optional) ranges
- Full-scale inputs correspond to 150V, 5A, 500W, 500VARs
per phase. Other full scale ranges [100-155V, 1.5-5.5A,
100-550 W/VAR] programmable (at reduced accuracy). Other
full scale inputs available via special order.
- Frequency transducer full scale input programmable to
any frequency between between 50.00 - 70.00Hz, unipolar
or bipolar.
- 1, 2, or 3 element power transducers accepted. Special
scaling for 2.5 element transducers selected via menu (1mA
range only).
- Quick connect cable connector provided for channels #1-8,
five-way binding posts for channel #9
- Load on selected channel adjustable (70 ohms - 100 kohms
+/-1% +/-5ohms in 120 ohms steps.
|
|
%Error Measurement
|
- Error between device under test and internal reference
standard is displayed directly in percent of full scale
for volts, amps, Watts and VARs. [Bracketed values apply
to menu programmed non-standard output ranges]
- Range: 0.01%-9.99% [0.1%-9.9%]
- Accuracy: +/-0.1% [+/-0.2%] (volts, amps, Watts), or +/-0.125%
[+/-0.25%] VARs, for values of 10%-100% of full scale and
load resistance > 5% of maximum value.
- Rated accuracy maintained for load resistances 5-100%
of ( 10kohm (0-1mA range), 700ohm (4-20mA range) ).
- Frequency error measurement: +/-0.135% of full scale in
mA
- Minimum 20 minutes warmup time required to reach rated
accuracy, 30 minutes if initial instrument temperature is
less than 18 deg. C
- Alarm limit programmable 0.10%-2.00%
- External milliamp measurement accuracy: +/-0.125% of full
scale
|
|
RS-232C Interface
|
- Standard DB-25 female connector
- Baud rates: 300 to 19200 baud
- Easy-to-use terminal mode
- Interface is optically isolated from power circuitry
|
| Stepping: |
- Voltage, current, Watts, VARs, or frequency may be stepped
- Start, stop, step value, and delay between steps independently
adjustable
|
|
Configuration:
|
- Multiple configuration setups with custom procedure name,
voltage current output selection, reference input selection,
P.T., C.T. ratios, and user defined prompts can be saved
in non-volatile memory.
- Primary readings can be displayed
|
|
Default values:
|
- Power-up default values of all output parameters plus
primary P.T. C.T. ratios can be saved in non-volatile memory.
|
|
Physical Characteristics
|
- 14.5"W x 7.5"H x 15.5"D (36.8cm W x 19cm
H x 39.4cm D)
- Weight: 25.0 lbs (11 kg)
|
|
Options
|
- Option 02: Cordura carry case
- Option 03: Test leads- 6ft leads for
connecting the voltage and current outputs to the transducer(s)
under test
- Option 04A: 120V nominal AC line voltage
input
- Option 04B: 240V nominal AC line voltage
input
- Option 05A: 60Hz nominal AC frequency
- Option 05B: 50Hz nominal AC frequency
- Option 06: 4-20mA range
- Option 07: Extra manual
- Option 08: 1 Year extended warranty.
Additional year for a total of 2 years
- Full-Service Agreement Preventative
maintenance, calibration and service plan
Download full
MTS-1400 Series brochure 148kB PDF
format |
NOTE: All specifications are subject
to change
|
 |
 |